Setting or query
Setting syntax:
ILOG <number>
Examples:
ILOG
2.8
ILOG .1
ILOGIC
2
Query syntax:
ILOG?
Query response examples:
Discussion:
This command sets the logic supply current limit to the
value specified. The units are amperes for both the setting
and query versions. The range is 0.10
A
to 3.0
A
and the
resolution is 100 mA.
INEGATIVE
Type:
Setting or query
Setting syntax:
INEG <number>
Examples:
INEG 1.45
INEG 1
INEGATIVE
.3
Query syntax:
INEG?
Query response examples:
INEG 1.1
;
INEG
.750;
Discussion:
This command sets the negative supply current limit to
the absolute value specified. The units for the setting and
query versions are amperes. The range is 0.050 A to
0.750
A
(1.6 A at 15
V
and below) in the high power com-
partment and 0.050
A
to 0.40 A (0.75
A
at 15 V and below)
in the standard compartment. The resolution is 0.050
A.
INEGATIVE