Operational
Operational syntax:
INIY
Discussion:
This command changes instrument settings to the pow-
er-on state. These settings are shown in Table
3-3
in this
section.
Type:
Setting or query
Setting syntax:
IPOS <number>
Examples:
IPOS .45
IPOS 1
IPOSlTlVE 0.3
Query syntax:
IPOS?
Query response examples:
IPOS
1.1;
IPOS .750;
Discussion:
This command sets the positive supply current limit to
the value specified. Units for the setting and query versions
are in amperes. The range is
0.05
A
to 0.750
A
(1.6
A
at
15 V and below) in the high power compartment and 0.05
A
to 0.40
A
(0.75
A
at
15
V and below) in the standard
com-
partment. The resolution is 0.05
A.