EasyManuals Logo

Tektronix TCP300 Series Service Manual

Tektronix TCP300 Series
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
TCPA300 and TCPA400 Performance Verication
Table 9: TCPA30
0 and TCPA400 AC coupling measurements
TEKPROBE Interface Cable BNC coaxial cable
Amplier
Square
wave
generator
amplitude
Vertical
gain
setting
DC
coupled
amplier
output
Minimum
AC
coupled
amplier
output
Maximum
AC
coupled
amplier
output
Vertical
gain
setting
DC
coupled
amplier
output
Minimum
AC
coupled
amplier
output
Maximum
AC
coupled
amplier
output
TCPA300
0.5 Vp
200 mA/div
1.00 Ap-p 1.05 Ap-p 1.50 Ap-p
200 mV/div
1.00 Vp-p 1.05 Vp-p 1.50 Vp-p
TCPA400
0.5 Vp
200 A/div
1000 Ap-p 1050 Ap-p 1500 Ap-p
200 mV/div
1.00 Vp-p 1.05 Vp-p 1.50 Vp-p
10. Disconnect the Calibration Adapter from the amplier.
Degauss
This is a functional test of the degauss operation. (See Figure 5.)
1. Connect a current probe to the output of the amplier. Do not clamp the
current probe around any conductor, but make sure the jaws are locked shut.
2. Press the amplier PROBE DEGAUSS AUTOBALANCE button. When the
indicator turns green, the degauss operation has successfully completed.
3. Verify that there are no error codes. Error codes display on the Probe Status
LEDs when the AC and DC COUPLING LEDs alternately ash.
4. Record the results (pass/fail) on the test record.
Figure 5: Setup for degaussing the current probe
14 TCPA300/400 Ampliers and TCP300/400 Series Current Probes Service Manual

Table of Contents

Other manuals for Tektronix TCP300 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TCP300 Series and is the answer not in the manual?

Tektronix TCP300 Series Specifications

General IconGeneral
BrandTektronix
ModelTCP300 Series
CategoryAmplifier
LanguageEnglish

Related product manuals