TCP303 and TCP404XL Performance Verication
TCP303 Curren
t Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP303 test record
Instrument S erial Number: Certicate Number:
Temperature
:
Relative Humidity %:
Date of Calibration:
Technician
:
Performance test Range, A/V Test Minimum Measured/calculated Maximum
1
-3% +3%
3
-3% +3%
5A/V
5
-3% +3%
0.6
-3% +3%
1.8
-3% +3%
DC Gain Accuracy
1
(% Error)
50 A/V
3.0
-3% +3%
Rise T
ime
5A/V
——
23 ns
Bandw
idth
5A/V
.707
1
DCGainAccuracyistestedatthe1Cto5Cspecication. If you need to test the probe at 0 °C to 10 °C, refer to the specication limits in the TCPA300/400
Ampl
iers & TCP300/400 Probes User Manual.
42 TCPA300/400 Ampliers and TCP300/400 Series Current Probes Service Manual