Command Descriptions
2-86
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
DATa?
INIT reinitializes the waveform data settings to their factory
defaults.
DATa INIT
Reinitializes the waveform data settings to their factory defaults:
DESTINATION = REFA (=TARGET)
ENCDG = RIBINARY
SOUrce = CH1
START = 1
STOP = 2500
WIDTH = 1
DATa?
Might return the following string:
:DATA:ENCDG RPBINARY;DESTINATION REFA; SOURCE REFB;START
1;STOP 500;WIDTH 2
CURVe, WAVFrm
DATa:DESTination
Sets or queries the reference memory location for storing oscillo-
scope waveform data that is transferred into the oscilloscope by the
CURVe command. This command is identical to the DATa:TARget
command.
Waveform
DATa:DESTination REF<x>
Arguments
Examples