Command Descriptions
2-92
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
CURVe?
DATa:TARget
Sets or queries the location for storing waveform data transferred
from an external device to the oscilloscope when executing the
CURVe command. This command is equivalent to the DATa:DES-
Tination command and is included here for compatibility with older
Tektronix oscilloscopes.
DATa:WIDth
Sets the number of bytes per waveform data point to be transferred
when executing the CURVe command. (Changing DATa:WIDth may
change t he following WFMPre parameters: BIT_Nr, BYT_Nr,
YMULt, YOFf, and YZEro.)
Waveform
DATa:WIDth <NR1>
DATa:WIDth?
<NR1> = 1 sets the number of bytes per waveform data point to 1
byte (8 bits).
<NR1> = 2 sets the number of bytes per waveform data point to
2 bytes (16 bits). If DATa:WIDth is set to 2, the least significant byte
is always zero.
DATa:WIDth 1
Sets the data width to 1 byte per data point for CURVe data.
Related Commands
Group
Syntax
Arguments
Examples