Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2-57
MULTICYcle displays a sine or square wave of several cycles. Default
for sine-like and square-like signals.
SINGLECYcle displays a sine or square wave of approximately one
cycle.
FFT displays the FFT of a sine wave.
RISING di splays the rising edge of a square wave.
FALLING displays the falling edge of a square wave.
FIELD displays a video signal synchronized on all fields. This is the
default for video signals.
ODD displays a video signal synchronized on odd fields.
EVEN displays a video signal synchronized on even fields.
LINE displays a video signal synchronized on all lines.
LINENum displays a video signal synchronized on the specified line
number.
DCLIne returns a query response when the oscilloscope finds a DC
level.
DEFault returns a query response when the oscilloscope cannot
determine the signal type.
NONE returns a query response when the AUTOSET menu is not
displayed. Set is ignored.
AUTOSet:VIEW EVEN
The oscilloscope will display video signals synchronized on even
fields when operated in autoset mode.
BUSY? (Query Only)
Returns the status of the oscilloscope. This command allows you to
synchronize the operation of the oscilloscope with your appli cation
program. Refer to Synchronization Methods on page 3--10 for more
information.
Arguments
Examples