iTRANS 2
FIXED POINT SINGLE OR DUAL GAS MONITOR
WITH DUAL ANALOG OUTPUTS
USER MANUAL
60
77036429-EN
Revision 7.0
Operation Commands
Operation commands are similar to write commands where specific values are written on specific
sensor to get the desired operation done. The details are enlisted in the
Table A - 5.
Command 150 Start/Abort Selected Live Channel Calibration Response/Request Length: 2
Byte
Number
Parsing Parameter
0 Unsigned-8 Selected Sensor (“1 = Sensor 1” and “2 = Sensor 2”)
1 Unsigned-8 Calibration Condition (“1 = Abort” and “2 = Start”)
Command 151 Start/Abort Selected Live Channel Zeroing Response/Request Length: 2
Byte
Number
Parsing Parameter
0 Unsigned-8 Selected Sensor (“1 = Sensor 1” and “2 = Sensor 2”)
1 Unsigned-8 Zeroing Condition (“1 = Abort” and “6 = Start”)
Table A - 5 Operation Commands