EasyManua.ls Logo

Teledyne 7600 - Figure 2-4: Internal Piping with Single or Dual Measuring Units

Teledyne 7600
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Model 7600
Teledyne Analytical Instruments 26
An internal piping diagram for an instrument with one or two
measuring units is shown in Figure 2-4. When there are two measuring
units, if a built-in oxygen sensor is used, it must be installed in
measuring unit 2. The diagram shows the combination of two cells used
in measuring unit #2. This is possible by combining ranges. When a
single measuring unit is used, there is only a single inlet and outlet gas
connection.
Figure 2-4: Internal Piping with Single or Dual Measuring Units
Depending on the options chosen, the analyzer can be configured in
several ways. Table 2-1 lists the possible configurations for single and
dual measuring units with 1 to 4 component meters.

Table of Contents