9 Performing the measurement
65
Enter measuring time (15 min). The required value can be entered in
the field.
Insert the flue gas probe into the exhaust pipe and position it in the hot
spot.
Start averaging with .
Stability time starts.
Press Continue to end the stability time manually.
After a maximum of 3 minutes, the stabilisation time ends and
the measurement starts automatically.
The system records the readings in the set measuring cycle.
During the measurement, the readings and the calculated
values are displayed.
Readings are automatically saved.
The measuring result is displayed after the measuring phase has
been completed.
9.5.6 CO undiluted
A multi-hole probe (0554 5762) must be connected.
Call up function: | CO undiluted
Start measurement: tap .
Reading is displayed.
End measurement: tap .