9 Performing the measurement
70
4
End measurement: tap .
9.5.11 Gas flow
The function is only available if the chosen fuel is a gas.
The gas burner capacity is calculated by means of the gas amount consumed.
To this end, a gas amount is input and its consumption read out at the gas
meter.
1
Call up function: | Gas Flow
2
Set the gas amount to watch for at the gas meter.
3
Set the calorific value of the burnt gas.
4
Start measurement: tap .
The test time is displayed.
5
When the set gas amount is reached: tap .
The calculated gas flow and the gas burner capacity (in KW) are
displayed.
Edit values
All values that can be modified have a dotted underlining.
9.5.12 Oil flow rate
The function is only available if the chosen fuel is an oil.
This function is used to calculate the capacity of the burner from the set oil
pressure and the oil flow rate of the oil nozzle.
1
Call up function: | Oil flow rate
2
Set oil flow rate of the oil nozzle and oil pressure.
The calculated oil burner capacity is displayed (in KW).