10 Maintenance
89
IMPORTANT
Leaking solvents and degreasers!
Damage to the instrument and sensors!
- Do not store solvents and degreasers, such as isopropanol, in the case.
IMPORTANT
Strong or harsh alcohol or brake cleaner!
Damage to the instrument!
- Do not use any strong or harsh alcohol or brake cleaner.
10.5 Drain condensate container
The fill level of the condensate container can be read from the markings. Hold
the instrument horizontally or vertically to check the fill level.
CAUTION
Weak mix of acids in the condensate!
Minor injuries!
- Avoid skin contact.
- Make sure that the condensate does not run over the housing.
CAUTION
Condensate entering the gas path!
Damage to sensors and flue gas pump!
- Do not empty condensate container while the flue gas pump is in
operation.