EasyManua.ls Logo

TESTO 300 LL NEXT LEVEL - Replace the probe module; Check;replace particle filter

TESTO 300 LL NEXT LEVEL
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Maintenance
95
2
Blow compressed air through flue gas
ducts of the probe module and probe
handle (see illustration). Do not use a
brush!
3
Fit probe module onto the probe handle and click into place.
10.9 Replace the probe module
Disconnect the flue gas probe from the measuring instrument.
1
Press the key on the top of the probe
handle and remove the probe module.
2
Fit new probe module and click into place.
10.10 Check/replace particle filter
Check particle filter
Particle filters of the modular flue gas probe must be checked regularly for
contamination: check visually by looking through the window of the filter
chamber.
If there is visible contamination or inadequate pump flow, replace the filter.
Replace particle filter
The filter chamber may contain condensate. This is not a malfunction
and will not cause incorrect readings.

Table of Contents

Related product manuals