Input basic programming data
32
Input basic programming data
Delayed reaction
During analysis, the response to any key presses may be delayed.
Select operating mode
Under the menu option "Operating mode" you can select the
type of analytical control. The Testomat
®
EVO TH CALgives you
scope to control timing or quantity via water meter or a combination
of both.
Smallest pause interval = 0 minutes between analyses. Longest
pause = 99 minutes.
The analysis interval (interval between two analyses) is determined
by the configured purge cycle, the programmed pause time (interval)
and the duration of analysis. The duration of analysis is directly de-
pendent on the measurement value.
Select timing control
Select in the menu => Basic program=> Mode of
operation => Time controlled
Confirm the selection with "OK"
(An " * " appears at the end of the line)
(The factory default is "Time controlled" " * ")
Now enter the interval.
Select quantity control
Smallest interval = 1 liter, largest interval = 9999 liters. The analysis
is performed once the programmed quantity of water has passed.
Before the analysis, the line and measuring chamber are purged
(observe programmed purge cycles).
Select in the menu => Basic program => Mode of oper-
ation => Volume interval
Confirm the selection with "OK"
(An " * " appears at the end of the line)
Select in the menu => Basic program => Interval =>
Volume
Confirm the selection with "OK"
Input the corresponding flow volume in liters
Confirm the input with "OK"
Now select the water meter .