EasyManua.ls Logo

Testomat EVO TH CAL - Import Basic Programming Data; Password Protection; Password Input

Testomat EVO TH CAL
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Input basic programming data
52
Import basic programming data
Select in the menu
=> Basic program => Function SD-Card => Import
basic data
one of the files in the root directory of the SD card with the ending
"ini" using the arrow keys.
Import the files with "OK".
If an error message occurs during the import, the format of the data is
defective. This may occur if the file has been edited. In this case no
settings are changed. Use a simple text editor for editing (e.g. Word-
Pad), not a word processor, since otherwise the formatting may be
changed!
All device settings are imported except
- Language settings
- Operating hours counter
- Indicator fill level
- Password
Password protection
For data input and settings in the basic program, a four-digit pass-
word can be used. If you have forgotten your password, please con-
tact your Heyl Neomeris service partner or water supplier.
Password input
Press the "M" button
This brings you to the basic menu for "Basic program"
Confirm "Basic program" with "OK"
The cursor flashes in the "PW: 000field.
Use the cursor keys to input the sequence of digits and confirm
with "OK".
You then see the selection menu for the basic programming.
No export of passwords
Please note that in Exporting settings, the password is not saved!
FUNCTION SD CARD
Store measurement
Store error -
Import basic data
Export basic data
NOTE
BASIC PROGRAM
PW: 0000

Table of Contents