Error messages/troubleshooting
61
Error number/error message
Description, possible
causes
Solution, troubleshooting measures
Prior power failure (also
when switching off)
Failure of the internal
24V power supply
Replace 24V fuse
For all other defects, the device must be
repaired
Connection to clock
broken
Device should be repaired, so that the clock
works. If the time is not relevant, you can
keep using the device.
Device was switched off
and backup battery for
clock is empty
Lithium battery is used up. Replace battery.
The function to store
measurement values
and/or error was activat-
ed, but no SD card was
inserted
Card is write-protected
The write-protect tab at the side of the SD
card must not be set to "LOCK"
Card not formatted or
formatted with an invalid
file system
Format the SD card with a FAT or FAT32 file
system
Error when accessing
SD card, because the
file is write-protected or
the card is full or defec-
tive.
Remove the write protection of the file
Wipe the card
Insert new card
The measurement scope
is exceeded
Select other indicator type (basic program)
The maintenance inter-
val was exceeded
Dosing pump is defec-
tive
No dosing message
from dosing pump
Check the cable to the dosing pump to en-
sure a correct connection
Replace dosing pump
Error in the optical unit
(light source or receiver
defective)
Swap measuring chamber mounting
The water is too cloudy /
soiled
Install an upstream water filter
Viewing windows are
soiled