Software Architecture
16
Smart-Vue Administrator Manual Thermo Fisher Scientific
Enable 21CFR Part 11 Use this checkbox to enable 21CFR
Part 11 compliance for the system. If
the checkbox is not checked, 21CFR
Part 11 compliance is not enabled.
Lock Interval (min) This field specifies the period of time
after which, if no activity occurs on
the computer, Smart-Vue Client will
lock and require you to re-enter your
password.
If this option is changed on the Smart-Vue server (enabled or disabled),
you must close all Smart-Vue Client applications and restart them for the
change to take effect. Enabling this option slightly modifies Smart-Vue
Client operation. Examples include requiring justification for making
changes to sensor status, making changes to sensor settings, the inability
to make changes to the software protection feature, and more.
A colored icon in the Settings menu bar in Smart-Vue Client indicates
whether this option is currently enabled or disabled on the server.
Option disabled on the server (blue)
Option enabled on the server (orange)
Figure 7. Smart-Vue Client showing 21 CFR Part 11 status