Preventive Maintenance Guidelines
Thermo Fisher Scientific Smart-Vue Administrator Manual 29
2.1.2.5 Interference with other
electronic devices
Smart-Vue products use the license-free ISM band (Industrial Scientific
Medical). This frequency is reserved for remote monitoring-type
applications, and is different from that used for cellular telephones and
wireless land-line phones. There is therefore no risk of frequency
interference or disturbance with these types of devices. The wireless
system used in Smart-Vue products complies with current international
standards and regulations for each country in which they are sold.
Note : It is worth noting that radio-frequency emissions from Smart-Vue
end-point modules are 80 times weaker than a cell phone.
2.1.2.6 Placing Smart-Vue
end-point module for best
performance
For optimal operation, follow these recommendations when physically
placing your device:
Do not place the end-point module within 40 cm (16 in.) of another
end-point module.
Make sure the wireless end-point module is not placed on an
electrical conduit or cable tray (such as for computer network cables).
For best results, place the end-point module so that it faces the
general direction of the receiver antenna.
Keep about 20 cm (8 in.) of clear space around the end-point module.
An end-point module that is placed between two refrigerators of close
proximity may not communicate effectively.
Make sure all cables, if any, are firmly attached