3 Troubleshooting
3.1 Error codes Complete list
-991: Download from memory failed
Check that the radio module got datalogging enabled.
Ask for just one measurement
Restart datalogging
-992: NOT USED
-993: NOT USED
-994: Link between Smart-Vue Server and the Smart-Service
Firewall: TCP port used between Smart-Vue Server and Smart-
Service must be open (Port 1091, 1092,…)
Restart Smart-Service from the Smart-Vue Server interface
-995: Smart service failure
Restart Smart-Service from the Smart-Vue Server interface
-996: Radio receiver failed
USB receiver plugged in locally:
Restart Smart-Service, if failed; reboot computer.
Unplug receiver for at least one minute; check that the two
LEDs turn red for three seconds, plug receiver back into the
same USB port.
Try on another USB port, check if the driver modifies the com
port affected.