Preventive Maintenance Guidelines
Thermo Fisher Scientific Smart-Vue Administrator Manual 41
• 4: CO2 / Temperature
• 5: 4-20 mA
• 6: Dry contact
• 7: Not used
• 8: Not used
• 9: Not used
• A: PT100 sensor
• B to F: Not Used
SN-SN: serial number from 0000 to FFFF (65535 values)
For example: 051750610001 = first digital temperature sensor module
operating at 915 MHz, manufactured in 2010 and programmed with
bench #5.
These details are provided here for your information.
2.1.4.6 Color ring (shrink tube)
scheme
• Digital temperature sensor: white
• Differential pressure sensor: no ring (sensor is embedded)
• CO2/Temp module: green
• Regular PT100 sensor, typically used for -80°C ULT Freezers: blue
• LN2 tank PT100 sensor: yellow
• Oven PT100 sensor: red
2.1.4.7 Battery When evaluating battery life, the Smart-Vue System does not query a
end-point module to check the current battery strength. This method
actually consumes battery power and is incompatible with our goal to
provide extremely long battery life in Smart-Vue modules. Instead,
battery life is based on a count of wireless board activity.