Preventive Maintenance Guidelines
Thermo Fisher Scientific Smart-Vue Administrator Manual 41
4: CO2 / Temperature
5: 4-20 mA
6: Dry contact
7: Not used
8: Not used
9: Not used
A: PT100 sensor
B to F: Not Used
SN-SN: serial number from 0000 to FFFF (65535 values)
For example: 051750610001 = first digital temperature sensor module
operating at 915 MHz, manufactured in 2010 and programmed with
bench #5.
These details are provided here for your information.
2.1.4.6 Color ring (shrink tube)
scheme
Digital temperature sensor: white
Differential pressure sensor: no ring (sensor is embedded)
CO2/Temp module: green
Regular PT100 sensor, typically used for -80°C ULT Freezers: blue
LN2 tank PT100 sensor: yellow
Oven PT100 sensor: red
2.1.4.7 Battery When evaluating battery life, the Smart-Vue System does not query a
end-point module to check the current battery strength. This method
actually consumes battery power and is incompatible with our goal to
provide extremely long battery life in Smart-Vue modules. Instead,
battery life is based on a count of wireless board activity.

Other manuals for Thermo Scientific Smart-Vue

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Smart-Vue and is the answer not in the manual?

Thermo Scientific Smart-Vue Specifications

General IconGeneral
BrandThermo Scientific
ModelSmart-Vue
CategoryControl Unit
LanguageEnglish