Preventive Maintenance Guidelines
38
Smart-Vue Administrator Manual Thermo Fisher Scientific
For double-end-point modules:
HIGH VALUE 1 A measured reading is above the high
limit value (after time delay, if
configured) on the first sensor.
LOW VALUE 1 A measured reading is below the low
limit value (after time delay, if
configured) on the first sensor.
SENSOR 1 FAIL A measurement value for sensor 1 is
not detected.
HIGH VALUE 2 A measured reading is above the high
limit value (after time delay, if
configured) on the second sensor.
LOW VALUE 2 A measured reading is below the low
limit value (after time delay, if
configured) on the second sensor.
SENSOR 2 FAIL A measurement value for sensor 2 is
not detected.
For all end-point modules:
LOW BATT Indicates that battery power is equal
to or below a value of (10%).
2.1.4.3 Short-press On single end-point modules, a short-press on the button forces the
end-point module to read the current sensor measurement and refreshes
the display.
On double-end-point modules, such as temperature/humidity and
temperature/CO
2
,
• Press once to update the display with a fresh reading of sensor 1.
• Press again to update the display with a fresh reading of sensor 2.