Preventive Maintenance Guidelines
Thermo Fisher Scientific Smart-Vue Administrator Manual 37
Figure 17. Smart-Vue end-point module indicating alarm condition
Note : It is important to note that for alarms triggered when a limit value
has been exceeded, the triangular indicator remains displayed on the
Smart-Vue end-point module screen until the alarm is acknowledged via
the software. The triangle remains even if the sensor value returns to a
level within alarm limit boundaries.
Warnings
The end-point module displays a specific label for each alarm that is
detected. These labels appear on the bottom line of the LCD.
For single end-point modules:
HIGH VALUE A measured reading is above the high
limit value (after time delay, if
configured).
LOW VALUE A measured reading is below the
low limit value (after time delay, if
configured).
SENSOR FAIL No functional sensor is detected.
Alarm icon
Alarm description