Troubleshooting
Thermo Fisher Scientific Smart-Vue Administrator Manual 55
Serial modem plugged in locally:
Restart Smart-Service, if failed; reboot computer.
Unplug the power supply for at least one minute; check that
the two LEDs turn red for three seconds.
Receiver over TCP/IP network:
Network issues… Ping the receiver to see if it responds.
Check COM port settings (COM Port Redirector device
manager and Smart-Vue Server settings)
Unplug receiver power supply for at least one minute and
plug it back in.
-997: Probe error (see diagnostic flowchart in next section)
-998: Wireless transmission failure (see diagnostic flowchart in
next section)
-999: General error (see diagnostic flowchart in next section)
3.2 Diagnostic
flowcharts
This section provides diagnostic flowcharts to guide you through
troubleshooting three specific errors:
-997: Problem occurring between the sensor and the wireless
module.
-998: General communication error.
-999: General error when measurements are not transmitted or when
communication is not working properly.