C-Link Protocol Commands
Hardware Configuration
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual B-35
Send: set date 03-19-05
Receive: set date 03-19-05 ok
set default params
This command sets all the parameters to their default values. This does not
affect the factory-calibrated parameters.
Send: set default params
Receive: set default params ok
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example that follows saves the parameters to FLASH
memory.
Send: set save params
Receive: set save params ok
ozonator
This command reports the ozonator on or off. The example below reports
that the ozonator is on.
Send: ozonator
Receive: ozonator on
set ozonator onoff
These commands set the ozonator on or off. The example below sets the
ozonator off.
Send: set ozonator off
Receive: set ozonator off ok
time
This command reports the current time (24-hour time). The example that
follows reports that the internal time is 2:15:30 pm.
Send: time
Receive: time 14:15:30
set time hh:mm:ss
hh = hours
mm = minutes
ss = seconds