Operation
Alarms Menu
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-61
Bench Temperature The Bench Temperature screen displays the current bench temperature and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 5 to 50 °C. If the bench temperature reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
• In the Main Menu, choose Alarms > Bench Temp.
Min and Max Bench Temperature
Limits
The Minimum Bench Temperature alarm limit screen is used to change the
minimum bench temperature alarm limit. The minimum and maximum
bench temperature screens function the same way.
• In the Main Menu, choose Alarms > Bench Temp > Min or Max.
• Use and to increment or decrement the numeric value.
• Press to save set to value as actual value.
LAMP TEMPERATURE:
ACTUAL MIN: 50.0
o
C
SET MIN TO: 52.0
o
C ?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
BENCH TEMPERATURE:
ACTUAL 32.3
o
C
>MIN 15.0
o
C
MAX 40.0
o
C
RANGE AVG DIAGS ALARM