Operation
Alarms Menu
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-59
O
3
Lamp Temperature The O
3
Lamp Temperature screen displays the current ozonator lamp
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 65 to 75 °C. If the O
3
lamp temperature reading
goes beyond either the minimum or maximum alarm limit, an alarm is
activated. The word “ALARM” appears in the Run screen and in the Main
Menu.
In the Main Menu, choose Alarms > O
3
Lamp Temp.
Min and Max O
3
Lamp
Temperature Limits
The Minimum O
3
Lamp Temperature alarm limit screen is used to change
the minimum O
3
Lamp temperature alarm limit. The minimum and
maximum O
3
Lamp temperature screens function the same way.
In the Main Menu, choose Alarms > O
3
Lamp Temp > Min or Max.
Use and to increment or decrement the numeric value.
Press to save set to value as actual value.
ALARMS:
ALARMS DETECTED 0
>O3 LAMP TEMP OK
LAMP TEMP OK
BENCH TEMP OK
PRESSURE OK
FLOW A OK
FLOW B OK
ª
RANGE AVG DIAGS ALARM
INTENSITY A OK
INTENSITY B OK
O3 CONCENTRATION OK
MOTHERBOARD STATUS OK
INTERFACE STATUS OK
I/O EXP STATUS OK
O3 LAMP TEMPERATURE:
ACTUAL 68.8
o
C
>MIN 60.0
o
C
MAX 80.0
o
C
RANGE AVG DIAGS ALARM