C-Link Protocol Commands
Communications Configuration
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual B-39
Receive: set host name analyzer01 ok
instr name
This command reports the instrument name.
Send: instr name
Receive: instr name
o3 Analyzer
03 Analyzer
instrument id
This command reports the instrument id.
Send: instrument id
Receive: instrument id 49
set instrument id value
This command sets the instrument id to value, where value is a decimal
number between 0 and 127 inclusive.
Note Sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
Send: set instrument id 50
Receive: set instrument id 50 ok
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The example that follows shows that the instrument is in
the remote mode.
Send: mode
Receive: mode remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The example
that follows sets the instrument to the local mode.
Send: set mode local
Receive: set mode local ok
program no
This command reports the analyzer’s model information and program
version number, which will be dependant on the current version.