EasyManua.ls Logo

Thermo 49i - Page 44

Thermo 49i
276 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
Calibration Factors Menu
3-14 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
• In the Main Menu, choose Calibration Factors > O
3
Coef.
• Use and to increment or decrement the coefficient value.
• Press to save the new coefficient.
Reset User Calibration Default The Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
• In the Main Menu, choose Service > Calibration Factors > Reset User
Cal Defaults.
• Press to warn user and enable restore with .
• Use to reset the calibration pressure reference value when pressed
after .
O3 COEFFICIENT:
O3: 600
SET COEF TO: 1.000
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL: RESTORE DEFAULT CAL:
RESTORE
RESTORE
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM

Table of Contents