EasyManua.ls Logo

ThermoFisher Scientific HyPerforma DynaDrive - Page 35

ThermoFisher Scientific HyPerforma DynaDrive
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 | Replacing parts
Figure 3.46. Sliding plastic piece up
to tubing.
Figure 3.47. Screwing the last plastic
piece to the tubing.
Rewiring the load cells for vessels with an E-Box
Important: Wires are calibrated by Mettler Toledo, and cannot be tampered
with. Do NOT cut or adjust the length of the wires for any reason.
WARNING: Do not attempt to remove or replace any 50 or 500 L DynaDrive
S.U.B. load cell cables without the use of a professional electrician. Damage to
the equipment and personal injury may occur.
1. Use a 5/32 in. Allen wrench to remove the two bolts of the summing block plate from
the bottom of the cart, and the four bolts in the summing block from the summing
block plate (Figure 3.48).
Figure 3.48. Summing block plate and summing block to remove
from cart.
Summing block
Summing block plate
2. Connect the load cell wires to the summing block according to the wire schematic in
Figure 3.49.
Thermo Scientific 31DynaDrive S.U.B. Service Guide

Table of Contents