GENERAL INFORMATION ON USAGE AND
GENERAL INFORMATION ON USAGE AND
MAINTENANCE
ENVIRONMENTAL CONDITIONS FOR USE
Indoor use Altitude up to 2,000m
Pollution degree II Temperature range: 5-40°C
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% rela-
tive humidity at 40°C
STORAGE, USAGE PERIOD AND OTHERS
1. Environmental conditions for use
Indoor use Altitude up to 2,000m
Pollution degree II Temperature range: 5-40°C
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly 50% rela-
tive humidity at 40°C
2. When storing the instrument, ensure that the following conditions are met:
(1) The instrument should not be splashed with water.
(2) Store the instrument where air pressure, temperature, moisture, ventilation, sunlight,
dust, salt/sulfurous air, etc. do not give any negative side effect.
(3) Do not store or transport the instrument on a slope or uneven surface or in an area
where it is subject to vibrations or instability.
(4) Do not store the instrument where chemicals are stored or gas is generated.
3. Usage period
8 years from delivery providing regular maintenance is performed (according to the self-cer-
tification [TOPCON data]).
SAFETY DESIGNATIONS PER IEC 60601-1 STANDARD
• Type of protection against electric shocks: Class I equipment
Class I equipment does not depend on basic insulation only for protection against electric
shocks. It can also be earthed; therefore, the metal parts with which one comes into contact
do not become conductive if the basic insulation fails.
• Degree of protection against harmful ingress of water: IPx0
CL-300 has no protection against ingress of water. (The degree of protection against harmful
ingress of water defined in IEC 60529 is IPx0)
• Classification according to the method(s) of sterilization or disinfection recommended by the
manufacturer: not applicable.
• Classification according to the degree of safety of application in the presence of a flammable
anaesthetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in
the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide. CL-
300 should be used in environments where no flammable anesthetics and/or flammable gases
are presents.
• Classification according to the mode of operation: Continuos operation.
Continuos operation is the operation under normal load for an unlimited period, without the
specified limits of temperature being exceeded.