BEFORE REQUESTING SERVICE
BEFORE REQUESTING SERVICE
CAUTION MESSAGES
CHECK ITEMS
CLEAN THE COVER GLASS Carry out cleaning of the cover glass.
DIOPTER OVER
PRISM OVER
ERROR
Check to see that the lens is in the measurable scope.
Check to see that the lens is free from any flaw, dust or oil.
Clean both glasses and turn power on again.
PAPER END Printer paper is out. Load new paper.
CLOSE PRINTER COVER Close printer cover.
UV CALIBRATION NG Is the lens left on lens support?
Remove the lens from the lens support, and perform calibra-
tion again.
PRINTER HEAD OVER HEAT It is possible that the temperature of a printer head is rising.
Please turn off the power, and switch on a power supply
again after waiting for a while.
PRINTER CUTTER ERROR Please check whether the printer cover has closed or
whether there is not any foreign object in a printer.
PRINTER THERMISTOR NG Call service engineer.
INITIAL ERROR
(ERROR CODE)
Call service engineer. The error code of four digits is dis-
played.
The instrument does not get
ready for operation even if the
power switch is turned on.
Re-plug the power cord.
Operation of a button is not
effective.
Power supply cannot be turned
off even if you push the power
switch.
Press and hold the power switch for 10 seconds or more.
Then, the power supply turns off.
S, C values are wrong. Is the lens place with power on?
Remove the lens and turn on power again.
Is the measuring beam blocked by, dust, marks, grease, etc.
in the measured lens?
Marking is poor. Replace the marking ink cartridge. For a lens with sharp sur-
face curve, use the optional steel needle marking set.
The screen went out all of a
sudden.
The auto shut-off function is on.
Tap the control power, and the instrument will resume.
Pushed the PRINT button but
the printer does not work.
Is printer paper set properly?
Not inside out?