2
CAUTIONS FOR USE
Before using this instrument, read the instruction manual. Fully understand the cautions for safety and the
use method.
Caution for use (Don't apply the instrument to the following patients.)
Patients whose treatment tissues cannot be properly checked (cloudy cornea, cataract, cloudy crystal-
line lens)
Important basic cautions
Give a detailed explanation to the patient about the expected effects and harmful symptoms before
operation.
Tell the patient not to move his/her body (especially, head) carelessly during operation.
Handling
Don't use this instrument where there are a combustible anesthetic, oxygen gas or other mixed gases.
[Explosion may occur.]
When the instrument is under observation or other conditions except laser emission, set it in the
standby mode.
[It may emit the laser beam wrongly.]
Keep your place at the instrument during operation. If the operator must leave the instrument, set the
key switch to "OFF", remove the key from the instrument and store it in a proper place.
[Others except the operator may use the instrument to cause an accident.]
Before emitting the laser, make sure that there is no reflective matter in its optical path.
[The laser beam may be emitted in an unexpected direction.]
In the inspection before using the instrument, make sure that the optical axis of the aiming beam is fit to
that of the treatment beam.
[The laser beam may not be emitted correctly.]
Be careful not to damage the optical components and soil them with fingerprint, dust, etc.
[The laser emission function may be deteriorated.]
When moving the chinrest up and down, be careful not to catch the patient's hand.
[The patient may be injured.]
Cautions for transportation/movement
• In transportation and installation, be very careful not to give a hard vibration or shock to the instrument.
• Don't pull the power cable whether the instrument is moved or not.
Disposal
When you dispose of the instrument, follow the regulations of each local government about disposal/
recycling.
ENVIRONMENT FOR USE
Temperature: 10~40°C
Humidity : 30~85% (without dew condensation)
Air pressure : 700~1060 hPa
STORAGE, USE PERIOD AND OTHERS
1. Environmental conditions
Temperature: 10~40°C
Humidity : 30~85% (without dew condensation)
Air pressure : 700~1060 hPa
2. Use period
8 years since delivery only when the regular maintenance/inspection is performed [according to the
self-certification (data of TOPCON)]