34
BASIC OPERATIONS
BASIC OPERATIONS
* Before using, refer to descriptions about installation and make sure again that SL-10L G and other
accessories are connected correctly with the main body.
WARNING
Both the beam and diffused beam are dangerous. Do not look at/touch
these.
During operation, let the people around the instrument wear the gog-
gles (applied to 532nm, OD4 or more) suitable for the laser wave-
length. The laser beam may damage their eyes.
First, make sure that the slit lamp attachment for laser photocoagulator
is at the specified position and then turn on the power of the instru-
ment. If it is not at the specified position, it may cause exposure to
laser beams and resultant burns and loss of sight.
CAUTION
The laser beam is emitted from the laser beam outlet. Keep your face
off the outlet as the beam may burn your eyes and skin.
Use this instrument following the procedure described in the Manual.
Controls and/or adjustments by procedures other than those described
here may cause exposure to laser beams and resultant burns and loss
of sight.
Place the Footswitch in a position where it is not in the way. To avoid
stumbling/injury to your foot, do not step too deep into the switch.
MEMO
If there is any anomaly, press the Emergency switch and stop the operation
immediately.
Since the optical fiber is easily broken structurally, handle it with particular care
and do not bend it forcedly.
When not in use, the key of the Key switch should be removed and kept by the
responsible manager.
Do not attempt to disassemble/rebuild the instrument by yourself.
Before emitting the laser beam, make sure that the instrument detects the con-
nected delivery units correctly.
If it does not detect them correctly, do not emit the laser beam. The instrument
may emit the laser beam exceeding the set value and it is very dangerous.