BASIC OPERATIONS
LIST OF FACTORS TO BE ADJUSTED DURING USING
Setting items that can be adjusted when using LC-300G are shown below. Since it would endanger
the patient if the laser beam is emitted by setting the emission output and emission time for excessively
large values, surely set these for proper values for treatment.
• Emission output
• Emission time
• Emission interval
• Aiming beam output
• Spot size
DIFFERENCE IN EMISSION BY EMISSION INTERVALS SETTING
<Emitting operation by changing settings using Repeat button>
• Single mode setting
When the Emission switch is pressed, the treatment laser beam is emitted once for the set emission
time.
• Repeat mode setting
While the Emission switch is pressed, the treatment laser beam is emitted for the set emission time
at the set intervals.
<Emitting operation setting by using Emission output dial>
• Continuous emission setting
The treatment laser beam is emitted while the emission switch is pressed.
Emission time and interval settings are invalidated.
Setting method of continuous emission and display
Fully turn the Emission time dial clockwise. Continuous emission is set when the Time
display window shows "–––".
When continuous emission is set, repeat mode is invalidated.