4
SAFETY CAUTIONS
Icons Prevention item Page
Prohibition
The endoprobe is a sterilized disposable product. Do not use it if the
sterilizing outer package gets wet (even once) or damaged, or if the
sealing is damaged.
32
Prohibition
Do not use the endoprobe if its validity time has expired.
32
Prohibition
When using the endoprobe, do not rebuild it in any way.
32
Prohibition
Both the beam and diffused beam are dangerous. Do not look at/
touch these.
34, 38
Mandatory Action
During operation, let the people around the instrument wear the
goggles (applied to 532nm, OD4 or more) suitable for the laser
wavelength. The laser beam may damage their eyes.
34, 38,
74
Mandatory Action
First, make sure that the slit lamp attachment for laser photocoagula-
tor is at the specified position and then turn on the power of the
instrument. If it is not at the specified position, it may cause exposure
to laser beams and resultant burns and loss of sight.
34
Mandatory Action
Firmly fasten screws and avoid death/injury due to falling.
49
Mandatory Action
Fix the protect filter between microscope body and the binocular tube,
and protect the eye from laser beams.
49
Prohibition
Do not open covers to avoid the danger of electric shocks. Ask a
service engineer for repairs.
52
Prohibition
Repairs/adjustments should always be done by the service engineer.
Electric shocks due to careless handling may cause death or serious
burns, and exposure to laser emission may cause burns and loss of
sight.
52
Mandatory Action
When changing fuses, shut down the power supply and pull off the
power cable. Removing the fuse cover without removing the power
cable may cause electric shocks.
57, 59
Mandatory Action
Be sure to use the attached fuse. Using another fuse may lead to fire
in case of a failure.
57
Mandatory Action
Let the observers whose microscopes (for assistant) have no protect
filters wear goggles (applied to 532nm, OD4 or more) suitable for the
laser wavelength. The laser beam may damage their eyes.
74
WARNINGS