EziWeigh5i, EziWeigh6i, EziWeigh7i, WOW2, DAS WOW2
Issue 2, 7/2017 35
ChnGd (current load cell code has had its calibration settings changed from factory
defaults YES/NO if yes, probably span or sensitivity calibrated),
CELL (current load cell output signal in mV/V. This feature can be used to observe faults
with the load bars, or determine sensitivity for a certain change in weight. The full scale range of the
indicator’s input is roughly -0.5 to +3.5 mV/V. Faulty load bars will often be observed at one of these
extremes or drifting values).
uUPd (current cell code micro volts per divisionbelow 3uV/d electrical noise may start to
affect the stability and accuracy of the scale).
Set to factory defaults
Press until displays Fact.d .
Press .
Press or to display, cLr.SE (clear setups) or cLr.AL (clear setups and
data).
Press to reset to factory defaults.
Weight units
Press until displays unitS .
Press .
Press or to select Std (kg) or Lb (pounds).
Language
Press until displays. LAnG .
Press .
Press or to select . EnG (English), ESP (Spanish), FrA (French),
dEu (German), or Por (Portuguese) options.
Version
Press until displays UEr (firmware version number).
Press to view firmware version.