10 Q-Trak™ XP IAQ Monitor Model 7585
Providing Power to the Q-Trak™ XP
The Q-Trak™ XP monitor can be powered with provided lithium-ion
battery or with the provided A/C adapter.
Operating the Instrument with the AC Adapter
The instrument can be
operated with or without
a battery using the
A/C adapter.
To operate the instrument
with AC adapter:
1. Plug the AC adapter
into an electrical
outlet.
2. Plug the other end
into the instrument.
In general, TSI
®
recommends operating the Q-Trak™ XP monitor
with a battery even when powering the instrument with an AC
power supply. Having a battery installed shortens the warm-up
time needed for sensors to make accurate measurements.
For best results, power on the instrument for 10 minutes before
verifying gas sensor performance. This allows the instrument and
sensors to thermally stabilize and reduces the amount of
measurement drift during verification or calibration.