The General Settings page is
used to configure the following
functions:
• Device Name is used to
customize the name of the
instrument. This feature is
helpful when associating
data to a specific instrument.
NOTE: The device will
display up to 14 characters
of the device name in the
header of the main
Dashboard page.
• Language is used to select
the desired language shown
on the instrument.
• Time Constant is used to
adjust the averaging period
for the readings shown on
the display.
• Sound is used to enable or disable the beeper. A beep is
emitted at the end of a Sample Interval when logging data or
when a Notification/Alarm occurs.
• Density Correction enables or disable the application of density
correction to the measurement of CO
2
and VOC.
• Show Out-of-Range Readings enables or disables out-
of- range readings shown on the device display. An out-of-
range reading is a measured value that is either less than
a sensor’s Minimum Detection Limit or greater than the
upper recommended range. For example, a negative gas
concentration reading is considered an out-of-range
reading. It is recommended that this setting is Enabled
when performing a Bump Test.