Operation and Service Manual 33
Hydrogen Sulfide
(H
2
S)
Hydrogen
Sulfide (H
2
S)
50
Zero Air
Ammonia (NH
3
)
Ammonia (NH
3
)
100
Zero Air
Formaldehyde
(CH
2
O)
Formaldehyde
(CH
2
O)
10
Zero Air
Gas Calibration Procedure
To perform a gas calibration,
required accessories and
material include a calibration
cap, ZERO calibration gas,
SPAN calibration gas, gas
regulator, and tubing. The
gas regulator used to control
the flow should be capable of
providing 0.3 L/min.
Select Calibration from the
Main Menu and the Calibration
page will appear showing the
available sensors for calibration.
All sensors appear grayed out
until you select the sensor they
would like to calibrate.
Step 1 – Select a Gas Sensor for Calibration
Remove the sensor module cover.