Operation and Service Manual 69
Sensor error. Service or
replace <SensorType>
sensor
Cannot initialize sensor due to
corrupt sensor memory or faulty
electronics.
Battery is faulty. Please
replace battery.
Battery has been depleted to
unsafe level.
The battery is too cold to
run. Saving data and
shutting down…
Battery temperature has
dropped below -20°C and is
unsafe to operate.
The battery is too hot to run.
Saving data and shutting
down…
Battery temperature has exceeded
60°C and is unsafe to operate.
Battery too cold to charge
Battery temperature is between
-20°C and 0°C and cannot
safely charge.
Battery critically low. Plug
in charging cable. Device
will shut down in
<30 Second Timer>
Apply AC power and allow battery
to charge.
Battery low. Plug the power
cord into the instrument or
auto shutdown will
occur soon.
Battery has less than 15 minutes of
runtime remaining.
Fan error. Service sensor
module.
Sensor module fan is blocked
or faulty.
Temperature sensor error.
Service sensor module.
Temperature sensor is faulty.
Barometric pressure sensor
error. Service sensor
module.
Barometric Pressure sensor is faulty.
Measurement Value =
88888888
Cannot calculate measurement
value possibly due to erroneous
User Calibration, faulty sensor, or
sensor removed.