4.3 Program 3 (3 Waste)
For waste materials when the manufacturer recommends autoclaving at a
temperature of 12C with a drying cycle.
Nominal Parameters
Sterilization temp.: 121ºC (250ºF).
Sterilization time: 20 minutes.
Dry time: 5 minutes.
Operations Sequence
Heating phase; water enters the chamber and heats up until the
sterilization temperature is reached.
Sterilization phase; temperature is maintained constant at the preset
level for the sterilization time.
Slow exhaust (Ex. Mode =3); steam is slowly exhausted from the
chamber, until it reaches a temperature of 8C.
Drying phase; Chamber is heated for 5 minutes at reduced power.
220
kPa
TIME
PRESSURE (kPa)
TEMPERATURE
Ambient Pressure
and Temperature
t1
t2 t3
t1 =
Heating stage
t2 = Sterilization stage
t3 = Exhaust stage
t4 = Drying stage
= Pressure
= Temperature
t4