9.2 Draining the Reservoir
Caution!
Before starting, Make sure that the electric cord is disconnected and
there is no pressure in the autoclave.
1. Connect the supplied plastic tube and turn drain valve counter
clockwise (1 turn).
2. When the water reservoir is empty, turn drain valve clockwise to the
clockwise position. Remove the plastic tube.
3. Fill reservoir with distilled water until it reaches the base of the safety
device holder (for water quantity, refer to tables on paragraphs 2.9,
2.10).
4. Connect the electric cord to power source.
The autoclave is now ready for use.