At the sterilization and dry stage, the countdown time left
to completion of the stage is displayed in the interval
between the readouts of temperature and pressure, on the
lower row.
The format of the display is MM:SS (two digits for
minutes and two digits for seconds).
Examples
Example 1: Autoclave between processes, the program No.1
has been selected.
Example 2: The autoclave in the sterilization stage, program
No.3 is running. Time left to completion is 3
minutes and 14 seconds.
Example 3: The process failed due to temperature drop in the
sterilization stage in program No.2.
2 - Instruments EXH.
LOW TEMP. 178K
1 - Instruments ST.BY
35
C 100K
3 - WASTE STER
122ºC 03:14 212K
Selected program
Waste
Stage
Temperature
Timer