a damp cloth.
11. Fill the reservoir with distilled water or mineral free water only.
12. Press the manual water fill button and allow a small amount of water
(2-4 ounces) to fill chamber. Remove water from chamber.
13. The instrument is ready to use.
IMPORTANT:
DO NOT sterilize instruments during the cleaning process!!!
CAUTION:
Keep out of reach of children. Contains mildly acidic ingredients.
Avoid contact with the skin, eyes or clothing. Wash hands well after
touching the powder, in the case of eye contact flush with continuous
running water for at least 15 minutes. If irritation persists get medical
attention. If accidentally swallowed, do not induce vomiting, drink
large amounts of water and obtain medical attention. MSDS available
upon request.
For models 2540 use one packet of CHAMBER BRITE™.
For models 3150, 3850, 3870 use two packets of CHAMBER BRITE™.
For models 5050, 5075 use three packets of CHAMBER BRITE™.
Clean every 20 cycles or as needed.