Attention!
If the perfectMelt/perfectMelt SE is connected to 230 volt the following compulsory
specifications must strictly be adhered to:
Connecting and initially operating the perfectMelt/perfectMelt SE must solely be conducted by
an electrically skilled person (comparable to German DIN VDE 0105-100 (2009-10)).
The connection terminals of the external conductors of the unit must be jumpered for that purpose.
This must be carried out with the help of a power-rated transversal section according to the appropriate
standards in the given country.
At the customer’s site the fusing of the phase must be power-rated and according to the
rules of the given country.
The supply line 230 V 1~/N/PE must be connected in a way that it is locked against rotation.
(Either with a 230 V CEE plug or the connection is made directly in the sub distribution.)
Please note that the maximum connected load available is 3000 watt only!
It is strictly prohibited to connect the unit with a shock-proof plug (comparable to the German Schuko
plug) because it does not provide a lock against rotation of the phases.
Improper connection can cause life-threatening situations.
We assume no liability, when these safety measures have not strictly been adhered to.
Power input
max. power input
2300 watt // 2000 watt
max. power input
2100 watt // 2000 watt
max. power input
2300 watt // 4000 watt
max. power input
2100 watt // 4000 watt
max. power input
2300 watt // 6000 watt
max. power input
2300 watt // 8000 watt