EasyManua.ls Logo

Vaisala DMT152 - Limit Alarm

Vaisala DMT152
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
38 ___________________________________________________________________ M210865EN-H
where
c1
=
Channel 1 analog output mode. Available options are:
1 = 0 ... 20 mA
2 = 4 ... 20 mA
4 = 0 ... 5 V
5 = 0 ... 10 V
7 = I alarm (ON/OFF output using current)
8 = U alarm (ON/OFF output using voltage)
Modes 1 ...5 are normal measurement
modes, where
malfunction alarm is always enabled, and limit alarm and
purge status alarm
are disabled.
c2
=
Channel 2 analog output mode. Available options are:
1 = 0 ... 20 mA
2 = 4 ... 20 mA
4 = 0 ... 5 V
5 = 0 ... 10 V
6 = LED alarm
7 = I alarm
(ON/OFF output using current)
8 = U alarm (ON/OFF output using voltage)
Enable mode 6 for channel 2 i
f you are using the Vaisala
LED cable.
Example:
amode 2 1
Ch1 output : 4 ... 20 mA
Ch2 output : 0 ... 20 mA
Limit Alarm
The limit alarm is activated when the measured quantity goes below the
low limit or above the high limit. The alarm can be specified during
ordering for a specified parameter, or configured on the serial line using
the ALARM command.
When ordering a limit alarm, the alarm levels are set as follows:
- For current output, an active alarm is indicated by the current being
at the maximum of the scale (20 mA).
- For voltage output, an active alarm is indicated by the voltage being
at the maximum of the selected scale. For example, if the voltage
output scale is 0 … 5 V, an output of 5 V indicates the alarm is active.
Using the ALARM command, you can also set a hysteresis value to
prevent the repeated triggering of the alarm when the measurement
fluctuates around a limit value.
ALARM<cr>

Table of Contents

Related product manuals