Selects the unit type to output: 
m = metric units
 (for example, celcius) 
n = nonmetric units (for example, fahrenheit) 
Set Serial Interface Mode 
SMODE [xxx]<cr> 
 
In STOP mode: outputting only when command is issued, any command 
can be used. 
In RUN mode: outputting automatically, only command S can be used. 
In POLL mode: outputting only when command is issued, any command 
can be used after the line has been opened using the OPEN command. 
The polling mode can be used when more than one transmitter is 
connected to one serial bus. When set to POLL state, the transmitters 
communicate one at a time when the specific transmitter address is called 
on the serial line. See descriptions of the commands ADDR on page 42 
and OPEN on page 52. 
Examples: 
smode stop 
Output mode    : STOP 
 
smode run 
Output mode    : RUN 
 
smode poll 
Output mode    : POLL