Serial Line Output Commands 
Start Measurement Output 
Use the R command to start the continuous outputting of measurement 
values as an ASCII text string to the serial line. The format of the 
measurement message is set with the FORM command. 
R<cr> 
Example (measurement message in default format): 
r 
Tdf= -2.5 'C H2O=  4919 ppm 
Tdf= -2.5 'C H2O=  4918 ppm 
Tdf= -2.5 'C H2O=  4917 ppm 
… 
 
Outputting the results continues in intervals issued with the command 
INTV. You can stop the output by pressing the Esc key, or with the 
command S. Since the interface is half-duplex, you must enter the 
commands when the transmitter is not outputting.  
Stop Measurement Output 
You can stop the measurement output by pressing Esc, or with the S 
command: 
S<cr> 
Set Output Interval 
INTV [n xxx]<cr> 
time interval = 0 ... 255 
time unit = "S", "MIN", or "H" 
 
Example: 
intv 1 min 
Output interval:    1 MIN 
 
The shortest output interval (with n = 0) is approximately one second due 
to the internal measurement cycle.