Field Check Using DM70 
To check of the accuracy of the DMT152 in the field, you need the 
following equipment: 
-  A fully charged MI70 indicator (the measurement display that is 
included in the DM70 package) 
-  A calibrated reference probe that is suitable for the intended 
measurement environment and dewpoint range. You can use a DMP74 
probe from the DM70 for dewpoint down to -60 °C  
(-76 °F), or a second DMT152 for full measurement range. 
-  A connection cable between DMT152 and the MI70 indicator (Vaisala 
order code 219980). 
-  A reference environment with a stable dewpoint and temperature in a 
suitable range for both probes. The normal measurement environment 
of the transmitter can also be used, if it meets these criteria. 
NOTE 
In the procedure below, the readings of the probes are checked 
individually. You can also perform the field check so 
that you connect 
the reference probe simultaneously with the DMT152 you are checking. 
In this case, you need two of the connection cables if you are using a 
DMT152 also as the reference probe. Also, you must power one of the 
transmitters by other means tha
n the MI70. 
Perform the check as follows: 
1.  Place the DMT152 in the reference environment (unless it is 
already). You can perform the check even if the DMT152 is 
currently being powered and used, as long as there is no cable 
connected to connector II. 
2.  Connect the MI70 connection cable between the MI70 indicator 
and connector II of the DMT152. 
3.  Turn on the MI70 indicator. If the analog channels of your 
DMT152 are configured to use the current output, and if the 
DMT152 is powered by the MI70, you will see an error message 
that indicates the supply voltage is too low for running the actual 
analog output. This is expected behavior; press OK to acknowledge 
the error and continue the field check. 
 
0805-014   
Figure 23  Supply Voltage Too Low for mA Output Error