EasyManua.ls Logo

Vaisala HUMICAP HMW90 SERIES - Connecting Common AC Power Supply

Vaisala HUMICAP HMW90 SERIES
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
30 ___________________________________________________________________ M211399EN-F
Wiring HMW93
Recommended wiring for long cables:
1111-068
Figure 14 Wiring HMW93
3-wire connection with -Vs as common ground. Maximum cable
resistance is 2.5 (24V supply, 0 ... 10 V output, relay not used).
1111-069
Figure 15 Three-Wire Wiring for HMW93
-T -Vs+T +Vs
Power supply
18 ... 35 VDC
or 24 VAC ±20%
-RH +RH
V
V
Relay
max. 50 VDC
500 mA
R
L
= 10 k min.
-T -Vs+T +Vs
Power supply
18 ... 35 VDC
or 24 VAC ±20%
-RH +RH
V
V
Relay
R
L
= 10 k min.

Table of Contents

Related product manuals